• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752917CP
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/13/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported about the loose tip error which occurred twice in a day during calibration before the procedure (procedure type was unknown).The procedures were completed with the replaced tips from an fms (fluidics management system) pak.The samples were not available.There was no involvement of a patient.This is one of two reports.
 
Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation; therefore, the condition of the product could not be verified.A review of the device history record traceable to the possible lot numbers, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A sample was not received at the manufacturing site and the device history record review of the possible lot numbers provided indicated the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown; therefore, specific action with regards to this complaint cannot be taken.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16361224
MDR Text Key309470679
Report Number1644019-2023-00148
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00000000000000
UDI-Public00000000000000
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 05/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065752917CP
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CENTURION VISION SYS, ACCESSORY, ACTIVE SENTRY HP; CENTURION VISION SYSTEM, ACCESSORY, HANDPIECE; CUSTOM-PAK SURGICAL PROCEDURE PACK
-
-