U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. MISSION KIT; BIOPSY INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. MISSION KIT; BIOPSY INSTRUMENT Back to Search Results
Model Number 2016MSK
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/18/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during computed tomography guided lung biopsy procedure, the outer distal portion of the coaxial cannula broke off in the patient.It was further reported that a surgical cutdown procedure was performed to remove the broke part of the coaxial cannula.Reportedly the broken piece was removed.The current patient status was unknown.
 
Additional Manufacturer Narrative
The manufacturing location was selected as unknown due to system limitations.The manufacturing location for the mission product is bd-santo domingo.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiration date: 05/2025).
 
Additional Manufacturer Narrative
H10: the manufacturing location was selected as unknown due to system limitations.The manufacturing location for the mission product is bd-santo domingo.H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: received one mission disposable core biopsy kit, which included the mission disposable core biopsy instrument, trocar stylet, compatible coaxial, blunt stylet and 10mm adapter for the evaluation.Upon visual evaluation, the device appeared to have residue throughout, and the device (mission needle) was received in initial configuration with cannula at 00 mm position.It appeared there was a complete circumferential break to the compatible coaxial.No other visual anomalies were noted on the returned device.Due to the nature of the complaint functional testing was not performed.Therefore, the investigation for the reported break issue is confirmed as the compatible coaxial was noted to be broken during visual evaluation.A definitive root cause for the reported break issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during computed tomography guided lung biopsy procedure, the outer distal portion of the coaxial cannula allegedly broke off in the patient.It was further reported that a surgical cutdown procedure was performed to remove the broke part of the coaxial cannula.Reportedly the broken piece was removed.The current patient status was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MISSION KIT
Common Device Name
BIOPSY INSTRUMENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe 85281
Manufacturer (Section G)
UNKNOWN
BR  
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key16361603
Report Number2020394-2023-00063
Device Sequence Number6757342
Product Code KNW
UDI-Device Identifier00801741097096
UDI-Public(01)00801741097096
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K171953
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2016MSK
Device Catalogue Number2016MSK
Device Lot Number0001475412
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/18/2023
Supplement Date Received by Manufacturer03/14/2023
Initial Report FDA Received Date02/13/2023
Supplement Report FDA Received Date03/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexMale
Patient RaceWhite
-
-