• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC UNKNOWN PINK SCREW #4; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC UNKNOWN PINK SCREW #4; SCREW, FIXATION, BONE Back to Search Results
Model Number 58802718
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Loss of Range of Motion (2032); Nonunion/Delayed-union Bone Fracture (2369); Implant Pain (4561)
Date of Event 01/18/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that the patient presented with pain in the right foot and in the implant.The patient had hardware removed due to mechanical failure/malfunction of fusion (right foot).The initial diagnosis was pseudarthrosis after fusion or arthrodesis.The pre-operative diagnosis was right hallux limitus with non-union.Non-union hallux interphalangeal joint fusion.The patient underwent revision right first metatarsal phalangeal joint fusion.Right foot hardware removal.Right hallux arthroplasty.Calcaneal bone graft harvesting.
 
Additional Manufacturer Narrative
The reported event, alleged of issue, "non-union" could be confirmed based on the operative report of the revision surgery.The device inspection revealed the following: a total of eight (8) devices were returned for evaluation.The devices consisted of one (1) broken plate and seven (7) screws.The broken plate and three (3) of the screws were determined to belong to the variax 2 system.The evaluation of the broken plate and the three (3) screws were performed under complaint pi 3216285, 3216514, 3216515, 3216516.Visual examination of the other four (4) screws revealed the screws are characteristic of legacy wright medical cannulated screw system (css).Slight scratches can be seen on the screw head.Lot number and size are not visible.No significant damages that indicate the screw contributed to the reported non-union could be found.More detailed information about the complaint event must be available in order to determine the root cause of the complaint event.¿ a review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.¿ a review of the labeling did not indicate any abnormalities.¿ ¿indications of material, manufacturing, or design related problems were unable to be identified as the catalog number and lot number were not communicated.¿ ¿ if more information is provided, the case will be reassessed.
 
Event or Problem Description
It was reported that the patient presented with pain in the right foot and in the implant.The patient had hardware removed due to mechanical failure/malfunction of fusion( right foot).The initial diagnosis was pseudarthrosis after fusion or arthrodesis.The pre-operative diagnosis was right hallux limitus with non-union.Non-union hallux interphalangeal joint fusion.The patient underwent revision right first metatarsal phalangeal joint fusion.Right foot hardware removal.Right hallux arthroplasty.Calcaneal bone graft harvesting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN PINK SCREW #4
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16361642
Report Number3010667733-2023-00063
Device Sequence Number6814439
Product Code HWC
UDI-Device Identifier00840420113047
UDI-Public00840420113047
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K102429
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number58802718
Device Catalogue Number58802718
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/19/2023
Supplement Date Received by Manufacturer04/05/2023
Initial Report FDA Received Date02/13/2023
Supplement Report FDA Received Date05/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient SexMale
Patient EthnicityHispanic
-
-