• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LDR MÉDICAL ROI-C ANCHORING PLATE; ROI-C TITANIUM-COATED IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LDR MÉDICAL ROI-C ANCHORING PLATE; ROI-C TITANIUM-COATED IMPLANT SYSTEM Back to Search Results
Catalog Number MC1005T
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/17/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3004788213-2023-00009 through 3004788213-2023-00011.
 
Event or Problem Description
It was reported that the roi-c plates had a hard time passing through the implant holder and caused the cage to break.The cage was removed and replaced with a second cage, but it also broke while trying to advance the plates.This cage was also removed and the holder and plates were switched out.The second set of plates were successfully installed into the third cage using the second holder.An awl was not needed and the patient did not have hard bone.There were no patient impacts.This is report three of three for this event.
 
Additional Manufacturer Narrative
Corrections in d4: udi number.Additional information in d4: expiration date, h4, and h6: component, investigation type, findings, and conclusions.Device evaluation: product was not returned, and photos were not provided.Device evaluation unable to be completed.Root cause: root cause was unable to be determined.This event could possibly be attributed to inserting the plates at an angle or not using a starter awl.Dhr review: per dhr review, the part was likely conforming when it left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event or Problem Description
It was reported that the roi-c plates had a hard time passing through the implant holder and caused the cage to break.The cage was removed and replaced with a second cage, but it also broke while trying to advance the plates.This cage was also removed and the holder and plates were switched out.The second set of plates were successfully installed into the third cage using the second holder.An awl was not needed and the patient did not have hard bone.There were no patient impacts.This is report three of three for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROI-C ANCHORING PLATE
Common Device Name
ROI-C TITANIUM-COATED IMPLANT SYSTEM
Manufacturer (Section D)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR  10300
Manufacturer (Section G)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR   10300
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key16362563
Report Number3004788213-2023-00011
Device Sequence Number15268541
Product Code OVE
UDI-Device Identifier03662663016971
UDI-Public(01)03662663016971(17)270301(10)289294/5
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K151934
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberMC1005T
Device Lot Number289294/5
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2023
Supplement Date Received by Manufacturer01/18/2023
Initial Report FDA Received Date02/13/2023
Supplement Report FDA Received Date12/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/28/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-