Correction: d1, d2a, d2b, d4 (catalog#, gtin), g4(510(k)), h3, h6 (method code).The reported event, alleged of issue, "non-union" could be confirmed based on the operative report of the revision surgery.The device inspection revealed the following: a total of eight (8) devices were returned for evaluation.The devices consisted of one (1) broken plate and seven (7) screws.The broken plate and three (3) of the screws were determined to belong to the variax 2 system.The evaluation of the broken plate and the three (3) screws were performed under complaint pi (b)(4).Visual examination of the other four (4) screws revealed the screws are characteristic of legacy wright medical cannulated screw system (css).Scratches and signs of wear can be seen on the screw head.The lot number and size of the screw are not visible.No significant damages that indicate the screw contributed to the reported non-union could be found.More detailed information about the complaint event must be available in order to determine the root cause of the complaint event.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the catalog number and lot number were not communicated.If more information is provided, the case will be reassessed.
|