STRYKER SUSTAINABILITY SOLUTIONS PHOENIX; CATHETER, RECORDING, ELECTRODE, REPROCESSED Back to Search Results
Model Number 200574
Medical Device Problem Code Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device had a buzzing sound.There was no patient injury or medical intervention and extended procedure time reported was five minutes.These are commonly used devices that are readily available.
 
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon inspection of the received complaint device, the device did not meet the curve specification.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The most likely root cause is mishandling subsequent to distribution, including shipping/storage conditions or improper manipulation of the catheter.The instructions for use (ifu) state: do not exert excessive pressure during placement of catheter if unknown resistance is encountered.Do not attempt to use the reprocessed ep catheter prior to completely reading and understanding the directions for use.Avoid manual pre-bending of distal curve, as this may damage steering mechanism of steerable catheter the reported event will continue to be monitored through post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
CATHETER, RECORDING, ELECTRODE, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS PHOENIX
10232 s. 51st st.
phoenix AZ 85044
Manufacturer (Section G)
STRYKER SUSTAINABILITY SOLUTIONS PHOENIX
10232 s. 51st st.
phoenix AZ 85044
Manufacturer Contact
andy dobos
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key16362831
Report Number0002090040-2023-00001
Device Sequence Number12577474
Product Code NLH
UDI-Device Identifier00885825003241
UDI-Public00885825003241
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K012708
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/17/2024
Device Model Number200574
Device Catalogue Number200574RR
Device Lot Number4848945
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/20/2023
Initial Report FDA Received Date02/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-