U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ADVANTIO; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ADVANTIO; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number K062
Medical Device Problem Codes Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438)
Health Effect - Clinical Code Fall (1848)
Date of Event 09/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the non-boston scientific right ventricular (rv) lead exhibited noise and high, out-of-range pace impedance measurements of grater than 2500 ohms.A signal artifact monitor (sam) event occurred earlier than the reported noise and high impedance measurements.Boston scientific technical services (ts) was contacted for a review.It was noted by the field representative that the patient had experienced a fall that month as well, wondering if that could be correlated.Ts discussed the possible options.The device was turned off, surgically abandoned and a non-boston scientific device was added.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that there was no conclusive evidence that the observed out-of-range impedance measurements were due to a malfunction or an inadequate lead-to-device connection.However, intermittent, out-of-range shock lead impedance measurements with no conclusive evidence of a malfunction or an inadequate lead-to-device connection are likely the result of the low-energy test signal utilized for daily automatic or in-clinic commanded shock lead impedance testing.A software update was released in 2015 to further improve consistency of the impedance test results and provide the clinician with additional diagnostic tools, including programmable shock lead impedance alert limits.
 
Event or Problem Description
It was reported that the non-boston scientific right ventricular (rv) lead exhibited noise and high, out-of-range pace impedance measurements of grater than 2500 ohms.A signal artifact monitor (sam) event occurred earlier than the reported noise and high impedance measurements.Boston scientific technical services (ts) was contacted for a review.It was noted by the field representative that the patient had experienced a fall that month as well, wondering if that could be correlated.Ts discussed the possible options.The device was turned off, surgically abandoned and a non-boston scientific device was added.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTIO
Common Device Name
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16364042
Report Number2124215-2023-06297
Device Sequence Number20718688
Product Code LWP
UDI-Device Identifier00802526516658
UDI-Public00802526516658
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
N970003/S132
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/19/2014
Device Model NumberK062
Device Catalogue NumberK062
Device Lot Number112566
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/26/2023
Supplement Date Received by Manufacturer02/17/2023
Initial Report FDA Received Date02/13/2023
Supplement Report FDA Received Date02/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/01/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age78 YR
Patient SexFemale
-
-