• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 4.3MM DRILL BIT/QC/180MM; BIT, DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 4.3MM DRILL BIT/QC/180MM; BIT, DRILL Back to Search Results
Catalog Number 310.431
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: device available for evaluation: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in colombia as follows: it was reported that on (b)(6) 2023, the drill bits were dull during an orthopedic procedure.This report involves one 4.3mm drill bit/qc/180mm.This is report 1 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device from the photo.Visual analysis of the photo revealed that the drill bit ø4.3 l180 f/quick coup appears to have the cutting edges of the fluted tip blunt and worn.The observed condition was consistent as an end of life indicator for the device.No other issues were identified.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.After a visual inspection per procedure, it was determined that the reusable instrument device was blunt from repeated use and servicing; therefore, further investigation for the reported complaint device is not required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the drill bit ø4.3 l180 f/quick coup would have contributed to the complained issue.There is no indication that a design or manufacturing issue has caused the complaint condition.It was determined that the reusable instrument was worn from repeated use and servicing.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the drill bit ø4.3 l180 f/quick coup appears to have the cutting edges of the fluted tip blunt and worn.The observed condition was consistent as an end of life indicator for the device.No other issues were identified.After a visual inspection, it was determined that the reusable instrument device was worn from repeated use and servicing; therefore, further investigation for the reported complaint device is not required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the drill bit ø4.3 l180 f/quick coup would have contributed to the complained issue.There is no indication that a design or manufacturing issue has caused the complaint condition.It was determined that the reusable instrument was worn from repeated use and servicing.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot part #: 310.431, lot #: 2l80064, manufacturing site: werk bettlach, release to warehouse date: 13 dec 2018, expiration date:n/a.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.3MM DRILL BIT/QC/180MM
Common Device Name
BIT, DRILL
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BETTLACH (CH)
muracherstrasse 3
bettlach CO 2544
SZ   2544
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16364570
Report Number8030965-2023-01758
Device Sequence Number20718818
Product Code HTW
UDI-Device Identifier07611819775145
UDI-Public(01)07611819775145
Combination Product (Y/N)N
Initial Reporter CountryCO
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number310.431
Device Lot Number2L80064
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/22/2023
Supplement Date Received by Manufacturer02/20/2023
06/23/2023
Initial Report FDA Received Date02/13/2023
Supplement Report FDA Received Date03/09/2023
07/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DRILL BIT Ø3.2 L145/120 2FLUTE F/QUICK C; LCP DRILL BIT Ø4.3 W/STOP L221 2FLUTE F/
Patient SexMale
-
-