| Model Number |
795210 |
| Medical Device Problem Code |
Unintended System Motion (1430)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was evaluated by an engineer.The engineer adjusted the cable for the affiniti 70 ultrasound system's release handle which resolved the reported issue.The system was returned to service.No further issues were reported.
|
| |
|
Event or Problem Description
|
|
A customer reported that their affiniti 70 ultrasound system's articulation arm was not locking in place.The failure occurred outside of clinical use and no patient or user was harmed as a result of the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
An addendum report is being submitted to provide the date received by manufacturer with the philips¿ become aware date.The information is in the g3 field.
|
| |
|