Model Number D134901
Medical Device Problem Codes Entrapment of Device (1212); Device Contamination with Body Fluid (2317); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Rupture (2208)
Date of Event 01/16/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Initial reporter phone : (b)(6).If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Biosense webster manufacturer's reference number (b)(4) has three reports: (1) mfr # 2029046-2023-00277 for product code d134801 (thermocool® smart touch® sf bi-directional navigation catheter) (2) mfr # 2029046-2023-00279 for product code d134801 (thermocool® smart touch® sf bi-directional navigation catheter) (3) mfr # 2029046-2023-00281 for product code d134901 (lasso® nav eco catheter).
 
Event or Problem Description
It was reported that a patient underwent a cardiac ablation procedure with two thermocool® smart touch® sf bi-directional navigation catheters and a lasso® nav eco catheter and suffered cardiac tamponade requiring a pericardiocentesis.It was reported that after ablation was conducted several times after insertion to the patient¿s body, sensor error (106) occurred.Pericardial fluid was accumulated after the procedure, about 300 cc of drainage was performed.Also, the tricuspid regurgitation (tr) value after the procedure seemed to be high.The biosense webster representative had explained that the ring catheter is contraindicated for the ventricles to the physician, but the physician used the catheter at the physician¿s discretion.The tricuspid valve was caught during right ventricular outflow tract (rvot) mapping, and the catheter could be removed at that time, but some tissue was attached to the removed ring catheter.The tissue was sent for biopsy.Timing when complaints occurred was after the catheter was inserted into the patient¿s body, and after ablation was conducted several times, when effusion was confirmed after the procedure.The cable was changed but the issue continued.The issue was resolved by changing the catheter to another one.At the time when the patient left the room, tr value was high, the vital signs were stable, but the chordae had been destroyed.The patient had to take diuretics.No cardiac surgery has been planned.The patient was admitted to the icu.The physician's opinions on the relationship between the event and the product was that the timing of pericardial effusion is unknown.There were no abnormalities observed prior to use of the product.The patient outcome of the adverse event is improved.Valve damage occurred at mapping phase.It is unknown when cardiac tamponade occurred.The force sensor error 106 is not mdr reportable.The biological material found on the catheter is not mdr reportable.The entrapment of device is a mdr reportable product malfunction.Since the event (cardiac tamponade and cardiac valve rupture) is life threatening and it might result in permanent impairment of a body function or permanent damage to a body structure; or it could require medical or surgical intervention to prevent permanent impairment of a body function or permanent damage to a body structure, it is to be considered serious and mdr-reportable.
 
Additional Manufacturer Narrative
Since no device has been received for analysis, no product investigation can be performed, and the customer complaint cannot be confirmed.Section h3.Has been updated.A manufacturing record evaluation (mre) was performed for the finished device and no internal action related to the complaint was found during the review.Additionally, the manufactured date has been provided.Therefore, field h4.Device manufacture date has been populated.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LASSO® NAV ECO CATHETER
Common Device Name
CATHETER, ELECTRODE RECORDING OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key16368686
Report Number2029046-2023-00281
Device Sequence Number7258751
Product Code DRF
UDI-Device Identifier10846835009651
UDI-Public10846835009651
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K113213
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/25/2025
Device Model NumberD134901
Device Catalogue NumberD134901
Device Lot Number30827527L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/16/2023
Supplement Date Received by Manufacturer05/26/2023
Initial Report FDA Received Date02/13/2023
Supplement Report FDA Received Date06/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SOUNDSTAR ECO SMS 8F CATHETER.; THMCL SMTCH SF BID, TC, D-D.; THMCL SMTCH SF BID, TC, D-D.; UNK BRAND CABLE.; UNK BRAND CABLE.
Outcome Attributed to Adverse Event Other;
Patient Age68 YR
Patient SexFemale
-
-