| Model Number |
VCP790D |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Component code: g07002 - device not returned.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental will be sent.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Trade name - irgacare® , active ingredient(s) ¿ triclosan, dosage form ¿ suture/solid/parenteral , strength ¿ = 275 g/m.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.During the procedure, the suture was easily detached from the needle during interrupted suturing.It was a control release needle.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) date sent to the fda: 3/28/2023 this is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: d9, h3, h6 h3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.The returned sample revealed that it was received one winding former with a detached needle, a suture, and one needle suture that pertain to product code vcp790d.During the visual inspection of the detached needle, the swage and attachment area was noted with a light swage.The barrel hole was examined under magnification for suture remnant, and none was observed.The suture end was examined and there isn¿t sufficient impression at the insertion mark.In addition, the needle suture combination was examined, and light swage was observed on the needle.In addition, the functional test was performed using instron equipment and the pull force result was below the minimum requirements.Based on the information currently available, iight swage issue was identified as pull out during the investigation of the sample received.This product issue will be addressed through the quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.
|
| |
|
Search Alerts/Recalls
|
|
|