MAUDE Adverse Event Report: ALCON RESEARCH, LLC SILIKON 1000 OIL; FLUID, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC SILIKON 1000 OIL; FLUID, INTRAOCULAR Back to Search Results
Catalog Number 8065601187
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pulmonary Edema (2020)
Date of Event 01/19/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported after the patient underwent ophthalmic surgery, the patient developed acute pulmonary edema that say day night and was transferred to hospital.The patient was in a state of using extracorporeal membrane oxygenation temporarily, but now that the patient¿s condition had calmed down and was returned from the hospital.
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information received indicating that the patient had underlying condition like retinal detachment, ulcerous colitis, serratus tear, proliferative changes, and subfoveal choroidal hemorrhage after trauma, and the patient was seen one month after the trauma that caused the injury.Multiple other traumas, including rib fractures and auricular hematoma, had occurred in a short period of time over the past year or so.The current trauma was caused by a falling frying pan at home.
 
Additional Manufacturer Narrative
Product was enrolled in the formulation stability program in which it is tested according to this product's stability specifications on a set schedule.Raw material testing is performed per requirements and disposition based on established acceptance criteria.All compounding, pre-processing, filling, and packaging mbrs are subjected to 2 independent reviews.A single, normal level i iba ansi z1.4 for attributes is performed for every lot manufactured.The process includes a review of the following: all chemistry and microbial in-process and finished product results; environmental, utility, bioburden records; sanitization records.Root cause could not be determined.Potential root causes include: - solution quality issue ¿ unlikely, as chemistry and microbiology data was reviewed and was found to have met regulatory requirements prior to release.- consumer mishandling - no conclusion can be made regarding the contribution of consumer mishandling as this factor is outside the control of the manufacturing facility.- event related to consumer physiology ¿ no conclusion can be made regarding the contribution of unique consumer physiology as this factor is outside the control of the manufacturing facility.Surgical practices cannot be confirmed, (e.G.During injection, etc.).The product labeling for silikon 1000 provides indications for use, contraindications, warnings, precautions, and directions for use to ensure proper use of the product.Instructions also state, ¿all components for single use only¿.No lot code was reported or sample provided by the customer.No further action is possible at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILIKON 1000 OIL
Common Device Name
FLUID, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC
6201 south freeway
fort worth TX 76134
Manufacturer (Section G)
ALCON RESEARCH, LLC
6201 south freeway
fort worth TX 76134
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16370222
Report Number1610287-2023-00005
Device Sequence Number20680249
Product Code LWL
UDI-Device Identifier00380656011873
UDI-Public00380656011873
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P950008
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number8065601187
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/30/2023
Supplement Date Received by Manufacturer02/14/2023
05/01/2023
Initial Report FDA Received Date02/14/2023
Supplement Report FDA Received Date03/15/2023
05/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization; Other;
Patient Age37 YR
Patient SexMale
-
-