| Model Number |
H7493941820250 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The target lesion was located in circumflex artery.A 2.50 x 20mm synergy xd drug-eluting stent was selected to treat the lesion.However, during preparation, the proximal part of the shaft broke when taking the catheter out of the shell.The procedure was completed with another of the same device.No patient complications were reported and the patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: synergy xd mr us stent delivery system was returned for analysis.Examination of the crimped stent via scope found no issues with the stent.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of tip damage.A visual and tactile examination of the hypotube identified a break on the hypotube shaft located at 21.8 cm distal to the distal end of strain relief as well as multiple kinks.The shaft polymer extrusion including the distal and mid-shaft sections were examined via scope and tactile examination.No issues were noted.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The target lesion was located in circumflex artery.A 2.50 x 20mm synergy xd drug-eluting stent was selected to treat the lesion.However, during preparation, the proximal part of the shaft broke when taking the catheter out of the shell.The procedure was completed with another of the same device.No patient complications were reported and the patient was stable.
|
| |
|
Search Alerts/Recalls
|