• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH TFNA FENESTRATED HELICAL BLADE 90MM - STERILE; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH TFNA FENESTRATED HELICAL BLADE 90MM - STERILE; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Model Number 04.038.390S
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/20/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: additional procode: hsb.Complainant part is not expected to be returned for manufacturer review/investigation.Initial reporter is a synthes employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from japan reports an event as follows: it was reported that on (b)(6) 2022, the patient underwent the primary surgery with the tfna for the femoral trochanteric fracture.At the time of injury, it was p type, so that the elevator was used to change to type a.Then, it was fixed.At one week postoperatively, the patient was progressing well with the expected amount of sliding.Approximately three months after surgery on (b)(6) 2023, cut-out was confirmed under a type.Although blade insertion direction was slightly forward, the cause of cut-out is unknown.A revision surgery was scheduled to be performed.No further information is available.This report is for a tfna fenestrated helical blade 90mm ¿ sterile.This is report 2 of 4 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot.Manufacturing location: elmira / packaged, sterilized and released by: monument.Manufacturing date: 25-mar-2022.Expiration date: 01-mar-2032.Part number: 04.038.390s, tfna fenestrated helical blade 85mm -sterile.Lot number: 721p884 (sterile).Lot quantity: 48.Note: helical blade was manufactured by elmira; lot number 691p759 qty 48.Parts were packaged, sterilized and released by monument.Production order traveler met all inspection acceptance criteria.Packaging label log was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Device history review: this lot met all visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TFNA FENESTRATED HELICAL BLADE 90MM - STERILE
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16372062
Report Number8030965-2023-01799
Device Sequence Number12577563
Product Code KTT
UDI-Device Identifier10886982099259
UDI-Public(01)10886982099259
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160167
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number04.038.390S
Device Catalogue Number04.038.390S
Device Lot Number721P884
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/23/2023
Supplement Date Received by Manufacturer03/09/2023
Initial Report FDA Received Date02/14/2023
Supplement Report FDA Received Date03/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LOCKSCR Ø5 L36 F/NAILS TAN LIGHT GREEN.; TFNA END CAP EXTENS. 0 TAN.; TFNA FEM NAIL Ø12 R 125° L235 TIMO15.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-