| Model Number |
367285 |
| Medical Device Problem Code |
Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Oem manufacturer: the manufacturing location for this product is nipro.This site is an oem manufacturing site.Therefore, bd corporate headquarters in franklin lakes, nj has been listed and the franklin lakes fda registration number has been used for the manufacture report number.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® safety-lok¿ blood collection set there was a shield activation issue.The following information was provided by the initial reporter.The customer stated: "the collection device does not have safety lock.".
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary: material #: [367285].Lot/batch #: [2j0611].Bd received 200 samples for investigation.The samples were evaluated by visual examination and the indicated failure mode for missing safety shield with the incident lot was not observed.Additionally, 50 retention samples from bd inventory were evaluated by visual examination and no issues were observed relating to the reported failure as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode missing safety shield.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® safety-lok¿ blood collection set there was a shield activation issue.The following information was provided by the initial reporter.The customer stated: "the collection device does not have safety lock.".
|
| |
|