| Model Number |
RNS 320 |
| Medical Device Problem Codes |
Failure to Interrogate (1332); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
Fall (1848)
|
| Date of Event |
01/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation results - investigation in process.Neuropace confirmed a significant dent on the explanted neurostimulator can.
|
| |
|
Event or Problem Description
|
|
The patient reported that he was unable to interrogate his device after experiencing a fall.The neurostimulator was replaced on (b)(6) 2023.A significant dent was noted on the top can of the neurostimulator when it was explanted.The neurostimulator was returned for investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation results: testing confirmed an intermittent connection in the pca circuit.This pca was subjected to a severe impact transmitted from the device can through the internal frame.Most probably the impact to the pca cracked some number of solder ball connections between the sdbga and the pca or possibly traces in the pcb itself were compromised.The intermittent electrical performance of this pca is consistent with the existence of intermittent circuit connections.
|
| |
|
Event or Problem Description
|
|
Original report - the patient reported that he was unable to interrogate his device after experiencing a fall.The neurostimulator was replaced on 1/16/23.A significant dent was noted on the top can of the neurostimulator when it was explanted.The neurostimulator was returned for investigation.Investigation results: testing confirmed an intermittent connection in the pca circuit.This pca was subjected to a severe impact transmitted from the device can through the internal frame.Most probably the impact to the pca cracked some number of solder ball connections between the sdbga and the pca or possibly traces in the pcb itself were compromised.The intermittent electrical performance of this pca is consistent with the existence of intermittent circuit connections.
|
| |
|