• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. SFX SPI MCRYL+ UNI 27IN 3-0 SA PS1 PR MP; SUTURE, SURGICAL, ABSORBABLE, POLYGLCOLIC ACID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. SFX SPI MCRYL+ UNI 27IN 3-0 SA PS1 PR MP; SUTURE, SURGICAL, ABSORBABLE, POLYGLCOLIC ACID Back to Search Results
Model Number SXMP1B104
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.During use on the patient it was noted that the needle broke.They used three packages of the device and the needles broke during closer.It was stated that there might have been some tension on the first one but not the last two.Another box with different lot number was used.There were no adverse consequences for the patient.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Component code: g07002 - device not returned.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Trade name: irgacare.Active ingredient(s): triclosan.Dosage form: suture/solid/parenteral.Strength: 2360 g /m.Related reports: 2210968-2023-01092, 2210968-2023-01093, & 2210968-2023-01095.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 3/21/2023.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found.H3 investigational summary: the product was returned to ethicon for evaluation.The returned sample revealed that it was received one unopened sample that pertain to product code sxmp1b104.In order to evaluate the condition of the returned sample, the packet was opened, and the swage and attachment area was noted to be as expected.The tip and body of the needle were examined under magnification and no defects or needle breakage were found.In addition, the sample was tested by resistance test to the needle and met the requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: the product was returned to ethicon for evaluation.The returned sample revealed that it was received one unopened sample that pertain to product code sxmp1b104.In order to evaluate the condition of the returned sample, the packet was opened, and the swage and attachment area was noted to be as expected.The tip and body of the needle were examined under magnification and no defects or needle breakage were found.In addition, the sample was tested by resistance test to the needle and met the requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: the product was returned to ethicon for evaluation.The returned sample revealed that it was received one unopened sample that pertain to product code sxmp1b104.In order to evaluate the condition of the returned sample, the packet was opened, and the swage and attachment area was noted to be as expected.The tip and body of the needle were examined under magnification and no defects or needle breakage were found.In addition, the sample was tested by resistance test to the needle and met the requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Related reports: 2210968-2023-01092, 2210968-2023-01093, & 2210968-2023-01095.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SFX SPI MCRYL+ UNI 27IN 3-0 SA PS1 PR MP
Common Device Name
SUTURE, SURGICAL, ABSORBABLE, POLYGLCOLIC ACID
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km. 8.3
san lorenzo PR
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key16375511
Report Number2210968-2023-01094
Device Sequence Number15268745
Product Code GAM
UDI-Device Identifier10705031236677
UDI-Public10705031236677
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
N12159
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model NumberSXMP1B104
Device Catalogue NumberSXMP1B104
Device Lot NumberSJBABM
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2023
Supplement Date Received by Manufacturer02/28/2023
Initial Report FDA Received Date02/14/2023
Supplement Report FDA Received Date03/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-