Model Number BC0410C-ID-700
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/16/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Items returned: n/a.Visual analysis: a visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation, nor were any images of the device provided in place of a device return.Procedure and medical imaging was not provided for this investigation.Investigation findings: without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.Without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.Based on a review of the device¿s risk documentation, the reported event did not indicate there were any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform and analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Ifu review (additional information can be found in the ifu): potential complications potential complications include, but are not limited to: vessel or aneurysm perforation, vasospasm, hematoma at the site of entry, embolism, ischemia, intracerebral/intracranial hemorrhage, pseudoaneurysm, seizure, stroke, infection, vessel dissection, thrombus formation, and death.Directions for use attach a 2-way stopcock to a 1cc syringe filled with appropriate contrast solution.Prime the 2-way stopcock so that no air is present.Slowly inflate the balloon to the recommended volume to achieve the desired diameter as described in table 3.Warning: do not exceed the maximum recommended inflation volume as balloon rupture may occur.Warning: always inflate and deflate the balloon while visualizing under fluoroscopy to ensure patient safety.After inflation, lock the stopcock if desired.If desired, remove the guidewire from the balloon catheter and prepare per the respective diagnostic or therapeutic agent ifu(s) for delivery through the guidewire lumen.Warning: excessive pressure higher than 700psi (4826kpa, 47.6atm) may cause leakage or rupture of the guidewire lumen.When deflating balloon, use fluoroscopy to ensure complete deflation prior to removal.See table 1 for respective deflation times.After procedure is complete, slowly remove the balloon catheter and guidewire.The device was not returned to the manufacturer for analysis; therefore, the alleged product issue cannot be confirmed.If the device or additional information is received, microvention, inc., will submit a supplemental mdr report.
 
Event or Problem Description
It was reported that the balloon was prepared properly and burst upon inflation during the procedure.The procedure was completed by using other product.No patient harm or injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCEPTER C
Common Device Name
CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
38 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key16376613
Report Number2032493-2023-00571
Device Sequence Number12357734
Product Code MJN
UDI-Device Identifier00842429103388
UDI-Public(01)00842429103388(11)220510(17)250430(10)0000190310
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K113698
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model NumberBC0410C-ID-700
Device Catalogue NumberBC0410C
Device Lot Number0000190310
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2023
Initial Report FDA Received Date02/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-