|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g4, g7, h1, h2, h3, h6, h9, h10 complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned product found the outer carton exhibits crush damage.The outer sterile blister is cracked.The inner sterile blister exhibits no damage.The sterility has not been breached.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.The root cause of the reported event can be attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|