• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. Z.S.G.M. CUTTER 3:1 RATIO ( CAUTION: SHARP ); SKIN GRAFT MESHER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. Z.S.G.M. CUTTER 3:1 RATIO ( CAUTION: SHARP ); SKIN GRAFT MESHER Back to Search Results
Model Number N/A
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event is recorded by zimmer biomet under (b)(4).The product is in the process of being evaluated by zimmer biomet.Once the product investigation has been completed, a follow up/final report will be submitted.Multiple mdr reports were filled for this event, please see associated reports: mfr - 0001526350-2023-00171, mfr - 0001526350-2023-00172, mfr - 0001526350-2023-00173, mfr - 0001526350-2023-00174.
 
Event or Problem Description
It was reported that during evaluation the cutter failed the test cut with an incomplete cut.There was no patient involvement.No adverse events were reported as a result of this malfunction.Due diligence is complete.No additional information is available.
 
Additional Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the cutter failed the mesh test cut with an incomplete cut; however, the repair was not completed because the cutter was returned without repair and notification that the cutter does not meet the zimmer mesh test acceptance criteria.Review of the device history record identified no deviations or anomalies during manufacturing relating to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filled for this event, please see associated reports: mfr - 0001526350-2023-00171-1, mfr - 0001526350-2023-00172-1, mfr - 0001526350-2023-00173-1, mfr - 0001526350-2023-00174-1.
 
Event or Problem Description
No additional information is available.
 
Event or Problem Description
There is no additional event information available.
 
Additional Manufacturer Narrative
If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Z.S.G.M. CUTTER 3:1 RATIO ( CAUTION: SHARP )
Common Device Name
SKIN GRAFT MESHER
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key16379203
Report Number0001526350-2023-00175
Device Sequence Number5898447
Product Code FZW
UDI-Device Identifier00889024375468
UDI-Public(01)00889024375468(11)200821(10)64839807
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
PREAMEND
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00770303000
Device Lot Number64839807
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2023
Supplement Date Received by Manufacturer06/06/2023
02/02/2023
Initial Report FDA Received Date02/15/2023
Supplement Report FDA Received Date06/09/2023
01/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/27/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-