|
It was reported that, during conization, ¿relax pressure on blade¿ was displayed and the device became not to function during use.It was found that there is a rack on the blade.The number of remaining use of hpblue was 79.Gen11 was used.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
|
(b)(4).Batch #: x95z5m.Additional information was requested and the following was obtained: in the event description states: there is a rack on the blade.Did the active titanium blade break off? no further information will be available.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the device was returned with the distal tip of the blade broken off returned.The remaining blade portion was scratched and with evidence of contact with metal in or out of the operative field.This blade tip portion may have broken off of the device during transport to our analysis site.During functional testing on gen11, an alert screen was displayed.A probable cause for the device to stop activating and the gen11 to display an alert screen is blade damage.The device was disassembled to inspect the internal components and no anomalies were found.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples, or clips during the procedure, or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in activation issues such as failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damaged blade can result in a broken blade.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot/batch x95z5m , and no non-conformances were identified.
|