|
|
| Model Number |
Y739D |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested and the following was obtained: could you please clarify how many devices present the defect? please confirm.Could you please clarify if just one lot number were involved? if not please provide the other lot number related.Yes, just one lot number.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Component code: g07002 device not returned.A manufacturing record evaluation was performed for the finished device batch, and no non-conformances were identified.Related events captured via 2210968-2023-01127, and 2210968-2023-01128.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on (b)(6) 2023 and suture was used.The sutures were broken during interrupted suturing.There were no adverse consequences to the patient.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 3/22/2023.H6 component code: g07002 no device problem found.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that it was received thirteen unopened samples pertain to the product code y739d.As per the sampling plan, a visual inspection was performed on thirteen samples, and no defects were found on the packages.The packets were opened, and the swage and attachment area were noted as expected.The sutures were dispensed without problems and examined along of the strand and no issues related to breakage sutures or anomalies were observed during the evaluation.Also, a functional test was performed using instron equipment and tensile force was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2023-01126, 2210968-2023-01127, 2210968-2023-01128.
|
| |
|
Search Alerts/Recalls
|
|
|