ETHICON INC. MONOCRYL SUTURE 18"(45CM) 2-0 VIO; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Model Number Y739D
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/06/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2023 and suture was used.The sutures were broken during interrupted suturing.There were no adverse consequences to the patient.Additional information has been requested.
 
Additional Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested and the following was obtained: could you please clarify how many devices present the defect? please confirm.Could you please clarify if just one lot number were involved? if not please provide the other lot number related.Yes, just one lot number.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Component code: g07002 device not returned.A manufacturing record evaluation was performed for the finished device batch, and no non-conformances were identified.Related events captured via 2210968-2023-01126, and 2210968-2023-01128.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 3/22/2023.H6 component code: g07002 no device problem found.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that it was received thirteen unopened samples pertain to the product code y739d.As per the sampling plan, a visual inspection was performed on thirteen samples, and no defects were found on the packages.The packets were opened, and the swage and attachment area were noted as expected.The sutures were dispensed without problems and examined along of the strand and no issues related to breakage sutures or anomalies were observed during the evaluation.Also, a functional test was performed using instron equipment and tensile force was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2023-01126, 2210968-2023-01128.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONOCRYL SUTURE 18"(45CM) 2-0 VIO
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
ciudad juarez 32604
MX   32604
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key16381431
Report Number2210968-2023-01127
Device Sequence Number5759244
Product Code GAN
UDI-Device Identifier10705031059542
UDI-Public10705031059542
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K960653
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberY739D
Device Catalogue NumberY739D
Device Lot NumberSEMPXC
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2023
Supplement Date Received by Manufacturer03/09/2023
Initial Report FDA Received Date02/15/2023
Supplement Report FDA Received Date03/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-