Model Number 1650
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the batteries would not charge.Both batteries resulted in yellow indicators on the battery charger regardless of how long they were plugged in regardless of which slot they were plugged into.The batteries were exchanged. no patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Investigation of this event is pending and a supplemental report will be sent upon its completion.Concomitant medical products: product id: dtpa2qq crt-d, implanted: (b)(6) 2022.Product id: 429888 lead, implanted: (b)(6) 2017.Product id: 6935m55 lead, implanted: (b)(6) 2017.Product id: 5076-45 lead, implanted: (b)(6) 2017.Additional products: brand name: heartware ventricular assist system ¿ batter, model #: 1650, catalog #: 1650, expiration date: 31-mar-2021, serial #: (b)(4), udi #: (b)(4), device available for evaluation: no, mfg date: 17-mar-2020, labeled for single use: no.Patient ime code(s): e2403.Imf code(s): f26.Img code(s): g02002.Fda device code(s): a0705.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
A supplemental report is being submitted for analysis and investigation completion.Two (2) batteries (b)(6) were returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the devices in relation to the reported event.Failure analysis of the returned devices revealed that the batteries passed visual inspection.Functional testing revealed that the batteries were unable to charge or provide power due to an enabled zero-volt charge (zvchg) field-effect transistor (fet).The gas gauge is not programmed to use the zvchg fet, which indicates that the batteries unintentionally enabled the fet.The zvchg fet disables the charge and discharge fets, rendering the batteries inoperable.If the battery with the zvchg fet enabled remains connected to the battery charger, the battery charger led status indicator will flash red after 8 hours due to a charge time-out.Internal inspection did not reveal any anomalies.As a result, the reported event was confirmed.Based on an investigation conducted under capa pr00519785, the most likely root cause of the reported event has been attributed to the battery connector interface design which does not guarantee the desired connection angle in some use conditions.(b)(6) and (b)(6) are in scope of fa1265, which was initiated for batteries with electrical faults.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
A supplemental is being submitted for a correction.Updated h10.Additional products: (b)(6) h6: fda device code(s): a0705 h6: fda method code(s): b01 h6: fda results code(s): c02 h6: fda conclusion code(s): d01 investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key16381895
Report Number3007042319-2023-00991
Device Sequence Number6279247
Product Code DSQ
UDI-Device Identifier00888707006873
UDI-Public00888707006873
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse Practitioner
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/31/2021
Device Model Number1650
Device Catalogue Number1650
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2023
Supplement Date Received by Manufacturer04/24/2023
04/24/2023
Initial Report FDA Received Date02/15/2023
Supplement Report FDA Received Date04/24/2023
04/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/17/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1103 VAD
Patient Age78 YR
Patient SexFemale
Patient Weight62 KG
-
-