| Model Number |
IPN902070 |
| Medical Device Problem Code |
Fitting Problem (2183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.(associated report 9 681900-2023-00004).
|
| |
|
Event or Problem Description
|
|
Reported event: during a resuscitation, a 50cc luer-lock failed to attach to the lma.A new lma size 3 was used and was able to attach a 50cc luer-lock.After the procedure another device was tested with a luer-lock 50cc and this also failed to fit.
|
| |
|
Event or Problem Description
|
|
Reported event: during a resuscitation, a 50cc luer-lock failed to attach to the lma.A new lma size 3 was used and was able to attach a 50cc luer-lock.After the procedure another device was tested with a luer-lock 50cc and this also failed to fit.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).No actual sample received for further investigation.Based on the complaint description of "luer connector not fitting during test / use", reviewing of the device history of the packaging and assembly lot, reviewing the nc list, it was concluded that the complaint root cause is undetermined/unknown as the assembly or packaging process for the affected lot manufacturing process did not have any record on the similar issue.Further investigation shall be performed once the complaint sample was returned to kulim.
|
| |
|
Search Alerts/Recalls
|