SMITH & NEPHEW ORTHOPAEDICS AG PLUS SL MIA DBL OFFSET ADAPT. LT 60/25MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED Back to Search Results
Model Number 75004612
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/24/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference:(b)(4).
 
Event or Problem Description
It was reported that, during a thr surgery, the tip of broach handle of one (1) plus sl mia dbl offset adapt.Lt 60/25mm broke off while broaching.The procedure was resumed, without any delay, using a s+n back-up device.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
It was reported that, during a total hip replacement surgery, the tip of broach handle of one (1) plus sl mia dbl offset adapt.Lt 60/25mm broke off while broaching.The device intended for use in treatment was not returned for investigation.A product evaluation was not possible.A review of the production documentation did not detect any deviation that could have contributed to the reported failure mode.The review of historical complaints for the alleged device revealed no additional complaints reported for the same batch, and one additional similar complaint for the same product number over the past 12 months with similar failure mode.A review of past corrective actions was performed.No further escalation is required.A review of the risk management documentation verifies the failure mode, occurrence and severity of the reported issue.The performed investigation do not lead to an accurately determined cause.There is no indication that the reported devices failed to meet manufacturing specifications upon release for distribution.Due to insufficient information, it is not possible to speculate about factors which could have contributed to the reported event.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.There is no need for further actions because of the limited information provided.Nevertheless, smith + nephew will continue to monitor this device for similar issues.This complaint will be reopened should additional information or the device be received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLUS SL MIA DBL OFFSET ADAPT. LT 60/25MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16383549
Report Number9613369-2023-00116
Device Sequence Number5114152
Product Code LWJ
UDI-Device Identifier07611996085464
UDI-Public07611996085464
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75004612
Device Catalogue Number75004612
Device Lot Number1581
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2023
Supplement Date Received by Manufacturer03/22/2023
Initial Report FDA Received Date02/15/2023
Supplement Report FDA Received Date03/28/2023
Date Device Manufactured03/28/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-