| Model Number |
KSP14100E |
| Medical Device Problem Codes |
Fracture (1260); Unintended Movement (3026)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/23/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient had an alleged dissociation of the resurfacing femur and stem extension and the stem extension fractured.The patient will be undergoing a revision surgery on (b)(6) 2023.No additional information regarding this event has been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is in process.When the investigation is complete, a supplemental mdr will be submitted accordingly.The following mdr has been submitted for the same adverse event: -3013450937-2023-00026.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had an alleged dissociation of the resurfacing femur and stem extension and the boss of the resurfacing femur fractured.The patient underwent a revision surgery on (b)(6) 2023 to revise the implants to a compress system, custom adaptor, and distal femur implant.No additional information regarding this event has been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to include additional and corrected information.The investigation is complete.The root cause of the reported adverse event could not be definitively determined.The device was not returned for evaluation.The device history records and sterilization batch records were reviewed and no issues during manufacturing or sterilization were identified that would have contributed to this complaint.If any additional information is obtained, a supplemental mdr will be filed accordingly.The following sections were updated: h3: device evaluated by manufacturer updated to no.H6: type of investigation code updated to 3331: analysis of production records.H6: type of investigation code updated to 4114: device not returned.H6: type of investigation code updated to 4117: device discarded.H6: type of investigation code updated to 4110: trend analysis.H6: type of investigation code updated to 4109: historical data analysis.H6: investigation findings code updated to 3221: no findings available.H6: investigation conclusions code updated to 4315: cause not established.
|
| |
|
Search Alerts/Recalls
|