RICHARD WOLF GMBH BOA VISION; SENSOR-URETERORENOSCOPE 8.7FR WL 680MM Back to Search Results
Model Number 7355071
Medical Device Problem Codes Loss of or Failure to Bond (1068); Difficult to Remove (1528); Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
The user facility representative reported that sensor-ureterorenoscope 8.7fr wl 680mm, part no.7355071, serial no.(b)(4), deflection cover bunched up and the surgeon could not pull the scope through the access sheath and ureter.The primary surgeon needed to call a partner who was able to help remove the scope.The secondary urology (the specialist) didn't seem upset about the issues.I am not sure why the deflection cover bunched up.Additional information was received from the user facility representative on january 20, 2023.According to the received information, there was a delay of 45 minutes in the procedure and a follow-up procedure is necessary to complete the scheduled and planned treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BOA VISION
Common Device Name
SENSOR-URETERORENOSCOPE 8.7FR WL 680MM
Manufacturer (Section D)
RICHARD WOLF GMBH
pforzheimer strasse 32
d-75438 knittlingen, germany
GM 
MDR Report Key16385091
Report Number1418479-2023-00001
Device Sequence Number12033773
Product Code FGB
UDI-Device Identifier04055207054840
UDI-Public04055207054840
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/15/2023
Report Date (Section F) 02/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number7355071
Device Catalogue Number7355071
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date01/20/2023
Device Age32 MO
Event Location Hospital
Date Report to Manufacturer02/15/2023
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date02/16/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA02/15/2023
-
-