Model Number XIFNT511
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Osteolysis (2377); Swelling/ Edema (4577)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient has a history of osi placement.After uncovery, patient found to have pain and bone loss.Debridement and grafting attempted.Pain and mild edema after 6 months after grafting.Decision made to remove.Tooth 19.Implant removed; site grafted.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Patient weight unknown / not provided.Date of event unknown / not provided.A summary investigation has been completed for bone loss and infection events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss and infection have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE¿ TAPERED CERTAIN¿ IMPLANT 5 X 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key16385197
Report Number0001038806-2023-00309
Device Sequence Number5605469
Product Code DZE
UDI-Device Identifier00844868022015
UDI-Public(01)00844868022015(17)260630(10)2021060513
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K130949
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date06/30/2026
Device Model NumberXIFNT511
Device Catalogue NumberXIFNT511
Device Lot Number2021060513
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2023
Initial Report FDA Received Date02/16/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age34 YR
Patient SexFemale
-
-