• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CORPORATION, ASHITAKA RADIFOCUS GLIDECATH HYDROPHILIC COATED CATH; CATHETER, INTRAVASCULAR, DIAGNOSTIC,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CORPORATION, ASHITAKA RADIFOCUS GLIDECATH HYDROPHILIC COATED CATH; CATHETER, INTRAVASCULAR, DIAGNOSTIC, Back to Search Results
Model Number N/A
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/25/2023
Type of Reportable Event Serious Injury
Event or Problem Description
The user facility reported that the doctor used the glidecath device involved (cg415) to deliver.035 coils during a type ii endo leak case.The device was placed up the patient's leg of a covered stent graft and into the aneurysm sack.Upon delivery of the device, an.035 azur coil item (45-751632) the doctor noticed that the coil was stuck and unable to advance or withdraw from the catheter.Upon additional effort the coil became detached from the coil wire.Upon further manipulation to try and deliver the coil and with draw the catheter the tip of the catheter, the device involved separated while the back portion came out.The patient was in stable condition and the case was able to proceed; however, a portion of the device was left in the patient.The procedure was able to be completed with a different catheter.There was no patient injury/medical or surgical intervention required.Additional information was received on 27 jan 2023: the approximate size of the catheter tip that was left in the patient was 5 centimeters.No additional intervention was planned to remove the catheter tip.The doctor decided to leave the catheter piece and proceed with coiling the area with a different catheter and additional coils.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIFOCUS GLIDECATH HYDROPHILIC COATED CATH
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC,
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya city, 418
JA  418
MDR Report Key16385434
Report Number2243441-2023-00004
Device Sequence Number20257456
Product Code DQO
UDI-Device Identifier04987350772190
UDI-Public04987350772190
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/16/2023
Report Date (Section F) 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model NumberN/A
Device Catalogue NumberCG415
Device Lot Number201116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date01/25/2023
Device Age26 MO
Event Location Hospital
Date Report to Manufacturer01/27/2023
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date02/16/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TERUMO AZUR COILS - ITEM 45-751632 ITEM 45-652050.
Outcome Attributed to Adverse Event Other;
Patient Age80 YR
Patient SexFemale
Date Report Sent to FDA02/16/2023
-
-