Model Number 16827
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Patient tracking received a tracking form through email reporting a catheter revision.The reason for revision was due to pump flipping.Additional communication with the reporter of the issue confirmed that the patient's pump was initially sutured down, but then began flipping.The reporter stated that the issue was diagnosed when the clinic was trying to perform a refill and the programmer wasn't reading the pump.The reporter stated that the device had to be manually flipped to perform the refill.The reporter stated that the pump is now sutured down to the muscle as opposed to the fat to ensure that the pump flipping would not happen again.No devices will be returned.
 
Additional Manufacturer Narrative
A review of the device history record, which includes verification of all steps in the manufacturing of the pump, verification of all final testing performed by/on the pump, verification of sterilization, and packaging for subject pump was performed.The review did not identify any non-conformances, issues or capas associated with pump function.Devices will not be returned.Per the instructions for use of the device, pump flipping is a known possible risk of use of the device.Internal complaint number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMETRA II PROGRAMMABLE PUMP
Common Device Name
IMPLANTABLE INFUSION PUMP
Manufacturer (Section D)
FLOWONIX MEDICAL, INC.
500 international drive
suite 200
mount olive NJ 07828
Manufacturer (Section G)
FLOWONIX MEDICAL, INC.
120 forbes blvd
suite 170
mansfield MA 02048
Manufacturer Contact
james bennett
500 international drive
suite 200
mount olive, NJ 07828
9734269229
MDR Report Key16385631
Report Number3010079947-2023-00021
Device Sequence Number5087339
Product Code LKK
UDI-Device Identifier00810335020242
UDI-Public(01)00810335020242(10)28007(17)220623
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P080012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/23/2022
Device Model Number16827
Device Catalogue Number16827
Device Lot Number28007
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2023
Initial Report FDA Received Date02/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-