| Model Number |
219557 |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The expiration date is currently unavailable.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
This is report 13 of 15 for (b)(4).It was reported by the field service engineer that during inspection, the customer observed that the restore fully fluted reamer 10.5mm (plus or minus 5 percent) device had contamination inside its cannulated section.According to the report, the event occurred before processing when the customer decided to inspect the device with their borescope.It was reported that the device was purchased on june 2022; and has never been used.It was further reported that the device was shelved and stored in controlled environment in the sterile processing department.There was no procedure nor patient involvement reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary : the complaint device has not been returned, therefore unavailable for a physical evaluation.However, photos were provided.Upon visual inspection of the photos, the devices were observed in a sterilization tray and not in its original packaging.Under magnification, one photo showed evidence of contamination in the canal.It is unknown which product was observed in the photo.This complaint can be confirmed.A manufacturing record evaluation was performed for the finished device 2203506, and no non-conformances were identified.The photo does not provide enough evidence to determine root cause.Physical evaluation should provide more information to discern a possible root cause.As per ifu, the precautions for this type of device consist of to inspect the condition of the device prior to use.For the instrument life, it is necessary to follow the instructions and warnings of any cleaning and disinfection agents and equipment used; also, the recommendation of the proper condition during the sterilization and maintenance to avoid any damage.Follow the instructions and warnings issued by the suppliers of any cleaning and disinfection agents and equipment used.Avoid exposing instruments to hypochlorite solutions, as these will promote corrosion.Avoid prolonged exposure to saline to minimize the chance of corrosion.Prior to sterilization, instruments should be cleaned by either automated or manual means described below.Dry the instrument immediately after final rinse.Use filtered compressed air to dry internal areas.After drying proceed to inspection instructions.Inspect all instruments before sterilization or storage to ensure the complete removal of soil from surfaces, tubes and holes, moveable parts.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the product was returned to mitek for evaluation.Mitek then conducted visual inspection of device received.Upon visual inspection, under magnification into the center canal, no rests of a brownish matter were observed.A manufacturing record evaluation was performed for the finished device lot: 2202001, and no non-conformances were identified.According with the physical inspection result, this complaint cannot be confirmed.The storage conditions after the device was manufactured is unknown.The device was manufactured on 3/1/2022 and is more than 1 year old.As per ifu, the precautions for this type of device consist of to inspect the condition of the device prior to use.For the instrument life, it is necessary to follow the instructions and warnings of any cleaning and disinfection agents and equipment used; also, the recommendation of the proper condition during the sterilization and maintenance to avoid any damage.Follow the instructions and warnings issued by the suppliers of any cleaning and disinfection agents and equipment used.Avoid exposing instruments to hypochlorite solutions, as these will promote corrosion.Avoid prolonged exposure to saline to minimize the chance of corrosion.Prior to sterilization, instruments should be cleaned by either automated or manual means described below.Dry the instrument immediately after final rinse.Use filtered compressed air to dry internal areas.After drying proceed to inspection instructions.Inspect all instruments before sterilization or storage to ensure the complete removal of soil from surfaces, tubes and holes, moveable parts.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.H10 correction narrative: the investigation has been updated to reflect the correct information.Investigation summary: the complaint device has not been returned, therefore unavailable for a physical evaluation.However, photos were provided.Upon visual inspection of the photos, the devices were observed in a sterilization tray and not in its original packaging.Under magnification, one photo showed evidence of contamination in the canal.It is unknown which product was observed in the photo.This complaint can be confirmed.A manufacturing record evaluation was performed for the finished device lot: 2202001, and no non-conformances were identified.The photo does not provide enough evidence to determine root cause.Physical evaluation should provide more information to discern a possible root cause.As per ifu, the precautions for this type of device consist of to inspect the condition of the device prior to use.For the instrument life, it is necessary to follow the instructions and warnings of any cleaning and disinfection agents and equipment used; also, the recommendation of the proper condition during the sterilization and maintenance to avoid any damage.Follow the instructions and warnings issued by the suppliers of any cleaning and disinfection agents and equipment used.Avoid exposing instruments to hypochlorite solutions, as these will promote corrosion.Avoid prolonged exposure to saline to minimize the chance of corrosion.Prior to sterilization, instruments should be cleaned by either automated or manual means described below.Dry the instrument immediately after final rinse.Use filtered compressed air to dry internal areas.After drying proceed to inspection instructions.Inspect all instruments before sterilization or storage to ensure the complete removal of soil from surfaces, tubes and holes, moveable parts.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
|
| |
|
Search Alerts/Recalls
|
|