IBAL UKA TIB BEA-RING TRL SZ 4 8MM |
| Health Effect - Clinical Codes |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2022, it was reported by a sales representative via sems that a ar-601-tbd8 tibial bearing trial is damaged.This occurred during a case, with no patient effect reported.During returned device evaluation, a reportable malfunction was discovered.
|
| |
|
Additional Manufacturer Narrative
|
|
During returned device evaluation, a reportable malfunction was discovered.Complaint is confirmed.One unpackaged ar-601-tbd8 serial/batch number (b)(4) was received for investigation.Visual inspection found that the ibalance uka tibial bearing trial had broken off at the elliptical profile.No fragments were returned for analysis.The root cause of the reported failure mode is undetermined; however, the observed event is most likely caused by excessive bending forces applied during use.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint is confirmed.One unpackaged ar-601-tbd8 serial/batch number: (b)(6) was received for investigation.Visual inspection found that the ibalance uka tibial bearing trial had broken off at the elliptical profile.No fragments were returned for analysis.The root cause of the reported failure mode is undetermined; however, the observed event is most likely caused by excessive bending forces applied during use.
|
| |
|