• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GN HEARING A/S GN RESOUND; HEARING AID, AIR-CONDUCTION WITH WIRELESS TECHNOLOGY,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GN HEARING A/S GN RESOUND; HEARING AID, AIR-CONDUCTION WITH WIRELESS TECHNOLOGY, Back to Search Results
Model Number 20472100
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer's ref#: (b)(4).Technical investigation concluded: a dhr review have been completed.No deviation or changes were found during manufacturing of the device.Trended case investigation concluded: the connector is broken, probably because the user has (not intentionally) misused it, while inserting the cable.If the cable has been slightly out of specification that could have had a negative effect on this.This has caused a low ohmic connection in the cable connector - almost a short circuit - that will emit heat when the cable is inserted into the wall charger, causing up to the rated 5 w of the wall charger to be dissipated inside the connector.Micro usb is a very well known and used standard but can mechanically break down.The clinical investigation concluded: a damaged power supply or a power supply of very poor quality could lead to burn injuries and in rare conditions there is a small risk that a wrong charging or discharging or a battery thermal runway can lead to high temperatures or fire accidents.The chargers have been tested according to applicable safety standards.These risks are reduced as far as possible and the probability of occurrence of these events is considered very low.There are no new clinical aspects (e.G.Unforeseen use or clinical conditions) discovered from this event.Therefore, it is concluded that the current clinical evaluation and risk/benefit conclusion herein are sufficient.Risk register reference: the residual risk after the risk control measure(s) is acceptable.No new hazardous situations revealed and no need to update the risk register.Manufacturer's investigation final.This is a combined initial and final report.
 
Event or Problem Description
On (b)(6) 2023 the charging port and cord was melted.User stated the charger was plugged in via the usb and not using the electrical plug.No harm to the user reported.Unknown if original accessories/equipment was used.No further information expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GN RESOUND
Common Device Name
HEARING AID, AIR-CONDUCTION WITH WIRELESS TECHNOLOGY,
Manufacturer (Section D)
GN HEARING A/S
lautrupbjerg 7
ballerup, 2750
DA  2750
Manufacturer Contact
karen frisk
lautrupbjerg 7
ballerup, 2750
DA   2750
MDR Report Key16388068
Report Number3005650109-2023-00022
Device Sequence Number5706337
Product Code OSM
UDI-Device Identifier05708296147113
UDI-Public05708296147113
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number20472100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2023
Initial Report FDA Received Date02/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/28/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexMale
-
-