SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS¿ SMARTSITE¿ GRAVITY SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 47233E
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/24/2023
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd alaris¿ smartsite¿ gravity set tubing was "stiff" and didn't connect well during use.The following information was provided by the initial reporter: "the tubing is now more stiff and not connecting well".
 
Manufacturer Narrative
H6: investigation summary: two samples were received and tested by our quality team.There was a drip chamber without attached tubing as well as a 447233e full set with the complaint of connection issues and stiff tubing.The intact tubing was tested and there were no issues or abnormalities after thorough examination.No root cause is available and it is assumed that this was the "stiff tubing" set.A device history record review for model 47233e lot number 22099386 was performed.There were no quality notifications issued for the failure mode reported by the customer during the build of this set.
 
Event Description
It was reported that the bd alaris¿ smartsite¿ gravity set tubing was "stiff" and didn't connect well during use.The following information was provided by the initial reporter: "the tubing is now more stiff and not connecting well".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS¿ SMARTSITE¿ GRAVITY SET
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16390332
MDR Text Key309743221
Report Number9616066-2023-00229
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10885403235443
UDI-Public10885403235443
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K820278
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number47233E
Device Lot Number22099386
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/18/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-