Model Number MRO-715A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus and the working channel issue was confirmed, the adapter doesn¿t fit the light post.This complaint is most likely the result of improper handling or maintenance of the device.During inspection and testing the following was also found: scratches on the window glass and debris under the lens, damaged light guide threads, bent and corroded outer tube.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event or Problem Description
The customer reported to olympus that during reprocessing a working channel issue was discovered, the scope light cord does not fit under the connector.There were no reports of patient harm associated with this event.
 
Additional Manufacturer Narrative
This supplemental report is to inform that upon further review, this is not a reportable malfunction.The initial medwatch reported the scope light cord does not fit under the connector.Per the legal manufacturer, there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.
 
Additional Manufacturer Narrative
This report is being submitted to correct the legal manufacturer¿s contact information and facility registration number.The facility registration number is 3011050570.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLIMLINE URETEROSCOPE,15CM
Common Device Name
URETEROSCOPE
Manufacturer (Section D)
GYRUS ACMI L.P.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI L.P.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16390693
Report Number3003790304-2023-00080
Device Sequence Number12033739
Product Code FGB
UDI-Device Identifier00821925010970
UDI-Public00821925010970
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K012758
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMRO-715A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2021
Supplement Date Received by Manufacturer02/24/2023
06/18/2024
Initial Report FDA Received Date02/16/2023
Supplement Report FDA Received Date02/24/2023
06/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/30/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-