• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. PS TIBIAL INSERTS SZ 2, 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. PS TIBIAL INSERTS SZ 2, 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 204-22-13
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Discomfort (2330); Ambulation or Postural Difficulties (2544); Balance Problems (4401); Breast Discomfort/Pain (4504); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/27/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported by the legal department that this female patient is being revised to a hinged knee.In recent months, patient has reported increase in knee pain.Surgeon thought it might be a chronic infection issue but believes this is a poly wear issue as imagine shows asymmetry in the poly.
 
Additional Manufacturer Narrative
Additional manufacturer narrative: pending evaluation.
 
Event or Problem Description
It was reported via legal documentation that a patient had a right knee revision and then experienced a second revision surgical procedure approximately 10 years after revision surgery.The patient experienced daily pain and discomfort in her right knee, which has limited her activities of daily living and impacted her qualify of life.Patient has suffered debilitating injuries and damages, including but not limited to, pain and discomfort; swelling; gait impairment; poor balance; component part loosening; soft tissue damage; bone loss; bone damage; and other injuries presently undiagnosed, which all require ongoing medical care.There was no other patient/medical information provided.No x-rays or images were provided.The device will not be returned.There is no other information available.
 
Additional Manufacturer Narrative
H10.Related: mfr #1038671-2024-02490 report 2 of 2.H11.Additional manufacturer narrative- additional information added.Report 1 of 2.D10.Concomitants: 200-02-32 - three peg patella 32mm, 2543858.204-01-02 - ps cemented femoral sz 2, 8221698.204-04-22 - trapezoid tibial tray sz 1f/2t, 2f/2t, 1200696.Pending investigation.
 
Additional Manufacturer Narrative
Multiple mdr reports were filed for this event, please see associated reports: 1038671-2024-02490.This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: h6 the following sections were corrected: b1 b2 g2 h6 the reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.The product associated with the reported event is within the scope of recall z-0019-2022 however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PS TIBIAL INSERTS SZ 2, 13MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gaiensville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key16395802
Report Number1038671-2023-00187
Device Sequence Number6384740
Product Code JWH
UDI-Device Identifier10885862048240
UDI-Public10885862048240
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K933610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/19/2017
Device Catalogue Number204-22-13
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/27/2023
Supplement Date Received by Manufacturer07/18/2024
01/29/2025
Initial Report FDA Received Date02/17/2023
Supplement Report FDA Received Date07/23/2024
01/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/19/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight159 KG
-
-