• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US CORAIL2 STD SIZE 12; CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US CORAIL2 STD SIZE 12; CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM Back to Search Results
Model Number 3L92512
Medical Device Problem Code Wrong Label (4073)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that a customer at (b)(6) made us aware that one of their corail stems (which should have a green label) has a blue label.We¿re not sure if this is a printer ink error, or why this is ¿ but it is incorrect & could cause an issue if it gets mistaken for a stem which should have a blue label.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned to depuy synthes for evaluation.The photo investigation revealed enough evidence to confirm the reported allegation.The colored marking is incorrect.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a search of the depuy nonconformance (nc) quality system found 1 unrelated past nc¿s associated with this product/lot combination.However an nc was raised to address the current complaint issue.1) quantity manufactured: (b)(6) parts 2) date of manufacture: 26 september 2022.3) any anomalies or deviations identified in dhr: 1 non-conformance associated with this lot, however there is no correlation between this non-conformance and the failure mode of the complaint.4) expiry date: 31 august 2027.5) ifu reference: w90918 ifu corail 2 hip stem.Dhr review retrieved from (b)(6), as the lot number was the same.Device history review: 1) quantity manufactured: (b)(6) parts 2) date of manufacture: 26 september 2022.3) any anomalies or deviations identified in dhr: 1 non-conformance associated with this lot, however there is no correlation between this non-conformance and the failure mode of the complaint.4) expiry date: 31 august 2027.5) ifu reference: w90918 ifu corail 2 hip stem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL2 STD SIZE 12
Common Device Name
CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key16395969
Report Number1818910-2023-03953
Device Sequence Number5544022
Product Code LZO
UDI-Device Identifier10603295168782
UDI-Public10603295168782
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K192946
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3L92512
Device Catalogue Number3L92512
Device Lot Number3927653
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2023
Supplement Date Received by Manufacturer07/17/2023
Initial Report FDA Received Date02/17/2023
Supplement Report FDA Received Date07/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-