Based on the investigation analysis, the reported event or mismatch at 8:44 am est on january 14, 2023 could not be confirmed as there was no corresponding calibration entry at the time.However, around the time of the event, the system displayed temporary mismatch between the sensor readings and fingerstick measurements.The reported event happened during the initial few days after sensor insertion, and during the initial settling period after sensor insertion, there could be chances of temporary mismatch, which would eventually normalize as the sensor stabilizes.Once the sensor stabilized after insertion, the system began performing within expectations and there was better accuracy between the sensor readings and fingerstick measurements.H3.Device evaluated by manufacturer? yes.H6.Investigation findings updated to 114.H6.Investigation conclusions updated to 4315.
|