| Catalog Number |
8065751762 |
| Medical Device Problem Code |
Operating System Becomes Nonfunctional (2996)
|
| Health Effect - Clinical Code |
Corneal Edema (1791)
|
| Date of Event |
02/08/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that during a cataract surgery while using an ophthalmic foot pedal an error message was displayed and pedal stopped working.Surgery was completed after taking new pedal and patient has experienced corneal edema and medication was given.
|
| |
|
Additional Manufacturer Narrative
|
|
The company representative did not confirm nor replicate anything the would have contributed to the reported event.There were no samples returned on this investigation.The system was tested and found to meet product specifications.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.Based on the information obtained, the root cause of the reported ¿system message and pedal not functioning¿ were inconclusive.The root cause of the reported ¿corneal edema¿ cannot be determined.Therefore, remains inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|