| Model Number |
27050E |
| Medical Device Problem Codes |
Temperature Problem (3022); Excessive Heating (4030)
|
| Health Effect - Clinical Codes |
Burn(s) (1757); Blister (4537)
|
| Date of Event |
12/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The event is filed under internal karl storz complaint id (b)(4).The evidence of a third-party repair to the shaft which is attached to the block was noted.The socket block was found to have oxidized metal inside, this socket block is attached to the shaft and where the high frequency cord is plugged into.The lot code of xy (10/201 is well over the warranty period).
|
| |
|
Event or Problem Description
|
|
It was reported that there was an event with a 27050e monopolar working element.According to the information received, doctor burned her hand when touching the casing (white plastic part on the working element) it was extremely hot.The shaft was not hot.The facility stated the doctor had a second degree burn with blisters.
|
| |
|
Additional Manufacturer Narrative
|
|
The event is filed under internal karl storz complaint id (b)(4).The cause of the damage is due to the user.In the teflon block are shimmering through black, this indicates a short circuit combustion inside.In addition, the contact is very corroded.The lock nut is broken.Since the article was already manufactured in october 2011 and it has a corresponding number of application behind it.0 similar complaints in the period of 01.01.2019 - 02.05.2023.
|
| |
|