• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMPLANT DIRECT SYBRON MANUFACTURING LLC LEGACY4 IMPLANT; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMPLANT DIRECT SYBRON MANUFACTURING LLC LEGACY4 IMPLANT; DENTAL IMPLANT Back to Search Results
Catalog Number 873716
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 01/10/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient weight is unknown.The implant failure modes, failure/ loss of osseo-integration and lack of primary stability are not product related and rather are attributed to patient contraindications, conditions, or clinician error in surgical protocol.No product investigation or corrective actions required.Complaints will continue to be trended.
 
Event or Problem Description
Per complaint (b)(4), during clinical procedure, patient experienced lack of primary stability.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGACY4 IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
IMPLANT DIRECT SYBRON MANUFACTURING LLC
3050 east hillcrest drive
thousand oaks CA 91362
Manufacturer Contact
sergio alvarado
3050 east hillcrest drive
thousand oaks, CA 91362
8057785975
MDR Report Key16397396
Report Number3001617766-2023-00609
Device Sequence Number17792762
Product Code DZE
UDI-Device Identifier10841307102885
UDI-Public10841307102885
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K192221
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Dentist
Remedial Action Inspection
Type of Report Initial
Report Date (Section B) 02/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number873716
Device Lot Number207231
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/31/2023
Initial Report FDA Received Date02/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexMale
-
-