| Catalog Number |
03183793122 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one patient sample tested with phos2 phosphate (inorganic) ver.2 on a cobas 6000 c (501) module (serial number (b)(4) ).No questionable results were reported outside of the laboratory.The sample initially resulted in a phosphate value of 1.0 mg/dl with a data flag.The sample was repeated, resulting in a value of 3.2 mg/dl and this value was deemed correct and reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
The calibration performed on 29-jan-2023 was ok and there were no alarms.The same reagent pack was calibrated on 30-jan-2023 and there were variations in the calibration signals.One level of quality control shifted outside of range on 29-jan-2023 and a second level of control shifted outside of range on 30-jan-2023.The field service engineer checked the system and could not find anything wrong.The customer replaced the reagent and no issues were observed.The customer believes the reagent "stayed out for a while".The investigation could not identify a product problem.The cause of the event could not be determined.
|
| |
|