| Model Number |
MC2LP-313S |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
Swelling/ Edema (4577)
|
| Date of Event |
12/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Our investigation is ongoing.A follow up report will be submitted when the investigation is complete.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Catheter showing loss of integrity (crack/crack) in the structure.The damaged part of the device was inside the patient, causing edema and leakage of administered solutions.
|
| |
|
Additional Manufacturer Narrative
|
|
No device was returned for evaluation and no photographs were provided.The contract manufacturer conducted a review of the manufacture records for the lot number reported.Their investigation revealed the device was manufactured and inspected according to specification with no non-conformances or abnormalities.The manufacture process includes a 100% leak test performed as a last step in the process.This test would have detected any leaks, holes, or weak spots if it had existed at the time of manufacture.The instructions for use (ifu) contains the following warnings and precautions: do not use sharp instruments near the extension tubing or catheter lumen.Care must be taken when using sharp objects or needles in close proximity to catheter lumen.Contact from sharp objects may cause catheter failure.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|