Model Number 5567-P-3013
Medical Device Problem Codes Component Missing (2306); Inaccurate Information (4051)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
T&e team received a femoral stem which was packaged in a box for a press-fit humeral stem.This event did not affect the outcome of the case.The stem box was labeled a shoulder reunion s stem but then inside the box was a hip stem.We realized this immediately and opened the correct size stem as we had backup implants for that case.As reported by rep, who was asked if the outer box was shrink wrapped, and if the stem was in an outer and inner sterile blister: "yes it was shrink wrapped.Stem was also in its sterile packaging correctly.Everything about the make up of the box and packaging was normal except for the stem inside being a hip stem instead of a reunion s shoulder stem as labeled.".
 
Event or Problem Description
T&e team received a femoral stem which was packaged in a box for a press-fit humeral stem.This event did not affect the outcome of the case.The stem box was labeled a shoulder reunion s stem but then inside the box was a hip stem.We realized this immediately and opened the correct size stem as we had backup implants for that case.As reported by rep, who was asked if the outer box was shrink wrapped, and if the stem was in an outer and inner sterile blister: "yes it was shrink wrapped.Stem was also in its sterile packaging correctly.Everything about the make up of the box and packaging was normal except for the stem inside being a hip stem instead of a reunion s shoulder stem as labeled.".
 
Additional Manufacturer Narrative
The reported event could be confirmed, since in the or, a rep from the t&e division found a hip stem from the jr division (cat: 1601-09132, lot: n97p27) in a package of a reunion s stem (cat: 5567-p-3013, lot: g7785558).The labeling on the box clearly indicated specifics of the t&e reunion s product, but the contents (implant) was wrong, and contained a femoral stem from jr.An nc was opened to investigate the event further.Based on investigation, the root cause was attributed to a manufacturing related issue.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESS-FIT HUMERAL STEM 13MM
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16400462
Report Number0008031020-2023-00090
Device Sequence Number5138467
Product Code KWS
UDI-Device Identifier07613327424973
UDI-Public07613327424973
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K202289
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5567-P-3013
Device Catalogue Number5567-P-3013
Device Lot NumberG7785558
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/26/2023
Supplement Date Received by Manufacturer03/20/2023
Initial Report FDA Received Date02/17/2023
Supplement Report FDA Received Date04/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-