• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH LOCKING SCREW,T8 FULL THREAD, 2.7MM/L20MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH LOCKING SCREW,T8 FULL THREAD, 2.7MM/L20MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 656320S
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
As reported: "the locking screw did not lock to the variax dr plate.No screw was used for that screw hole (the most distal, second screw hole from ulnar side).".
 
Additional Manufacturer Narrative
The reported event could be confirmed.The device inspection revealed the following: the identification of the returned locking screw was confirmed based on the catalog # and the lot # marked.The threads below the head, intended for locking, are heavily worn out and deformed.The wear observed is most probably due to excessive torque applied when inserting the screw and due to possible screw angulation outside of the range of -15° to 15°.The screw was tested with a fully functional variax distal radius plate.The screw kept spinning on the hole and the locking could not be performed as intended.Therefore, the functionality of the device could not be given and the reported event was confirmed.Based on investigation, the root cause was attributed to an user related issue.The screw could not be properly locked in the plate due to the damages observed on the screw's threads, most probably resulting from mishandling.Related to post market surveillance activities, a potential non-conformity report was initiated to address similar events related to the variaax2 locking feature.The deep investigation of the nc revealed that the application of over torque during insertion was the root cause of the event.This leads to the damage of the smart lock technology below the head.As a reminder, the operative technique clearly states that the proper use of the screw should prevent this deformation from happening: ¿caution ¿ with the use of variable speed power systems, the surgeon should initially reduce the power to the lowest setting.¿ final tightening of the screw should be performed by hand to avoid damaging the screw-plate interface.¿ a review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
 
Event or Problem Description
As reported: "the locking screw did not lock to the variax dr plate.No screw was used for that screw hole (the most distal, second screw hole from ulnar side).".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW,T8 FULL THREAD, 2.7MM/L20MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM   D-79111
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16400509
Report Number0008031020-2023-00089
Device Sequence Number5138481
Product Code HWC
UDI-Device Identifier07613327070712
UDI-Public07613327070712
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K180500
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number656320S
Device Catalogue Number656320S
Device Lot Number1000504198
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/26/2023
Supplement Date Received by Manufacturer03/02/2023
Initial Report FDA Received Date02/17/2023
Supplement Report FDA Received Date03/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-