• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P RSP SEMICONSTRAINED HUMERAL SOCKET INSERT, 40MM, HXE-PLUS; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P RSP SEMICONSTRAINED HUMERAL SOCKET INSERT, 40MM, HXE-PLUS; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Model Number 509-01-040
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 01/24/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Revision surgery due to infection.
 
Additional Manufacturer Narrative
See d4 expiration date, h4 and h5.The reason for this revision surgery, the agent reported "(infected patient 2 stage revision.Antibiotic spacer placed)".The previous surgery and the surgery detailed in this event occurred 10 months and 2 weeks apart.This evaluation is limited in scope as limited information was provided to djo surgical - austin for examination.If information regarding cultures identified in the infection, the severity of the infection or any other relevant information is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.First, there was a ncmr#: (b)(4) associated with the main part #: 510-08-000, rsp spacer w/retaining screw, which documents that out of 20 parts lot, 4 parts were rejected due to blast not uniform inside or scratches on blast.Later the rejected parts were reworked and accepted after proper justification.Second, there was a ncmr#: (b)(4) associated with the main part#: 508-40-101, glenoid, head w/retaining screw, rsp, 40mm, neutral, which documents that out of 15 parts lot, all parts were rejected due to tool marks on inside taper.Later the rejected parts were found with no defects and accepted after proper justification.All other items in the respective lot were met with the design, fit and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SEMICONSTRAINED HUMERAL SOCKET INSERT, 40MM, HXE-PLUS
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, TX 78758
MDR Report Key16401926
Report Number1644408-2023-00161
Device Sequence Number12166883
Product Code PHX
UDI-Device Identifier00888912144636
UDI-Public00888912144636
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K051075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/12/2026
Device Model Number509-01-040
Device Catalogue Number509-01-040
Device Lot Number383P1067
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2023
Supplement Date Received by Manufacturer02/17/2023
Initial Report FDA Received Date02/20/2023
Supplement Report FDA Received Date02/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
506-03-114 LOT: 831C2042.; 506-03-114 LOT:831C2021.; 506-03-130 LOT: 835C1877.; 506-03-134 LOT: 836C1397.; 508-32-204 LOT: 769P2244.; 508-40-101 LOT: 875C1315.; 510-08-000 LOT: 142G1361.; 530-10-108 LOT: 409T1708.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-