Model Number UHI-4
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was evaluated by olympus.The evaluation found that the allegation was confirmed.Additionally, due to a faulty main board component, the unit alarms and the front panel controls go blank.An additional issue with the rusted rear panel was identified during the device evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The customer reported to olympus that during preparation for use, the high flow insufflation unit intermittently shut down.No patient injury or procedure impact was reported.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that system shut down intermittently occurred due to the main board controlling the main function of uhi-4 failed to operate normally.The cause of the failure of the main board was not identified.In addition, it is likely that the alarm occurred due to a failure of the mainboard controlling the main function of uhi-4, and the front-panel control indication disappeared.The cause of the failure of the main board was not identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16402328
MDR Text Key310243167
Report Number3002808148-2023-01577
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeTC
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/21/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-