MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC DC EXTENSION CABLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC DC EXTENSION CABLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number C-VH-3030
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/23/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Trackwise id (b)(4).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event or Problem Description
The hospital reported that during preparation for an endoscopic vein harvesting procedure, dc extension cable rubber surrounding the cord was coming off.Damaged cord was not used during the case.A replacement device was used to complete the procedure.No patient involved.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Trackwise#: (b)(4).Updated sections: b4, g4, g7, h2, h3, h6, h10.The device was returned to the factory for evaluation on 02/14/2023.An investigation was conducted on 02/22/2023.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.The one end of the cable was observed to be intact, with no visual defects observed.The connector end that plugs into the device was observed to be pulled apart from the base of the connector, exposing the inner wires of the cable.No other visual defects were observed.Based on the returned condition of the device as well as the evaluation results, the reported failure "break" was confirmed.The reported device is an oem device, therefore, a lot history review was not applicable.The product is not returning.A lot number was not provided and the specific product lot number cannot be identified from a ship history, therefore it is impossible to obtain a certificate of conformance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DC EXTENSION CABLE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key16402908
Report Number2242352-2023-00126
Device Sequence Number14284620
Product Code GEI
UDI-Device Identifier00607567700888
UDI-Public00607567700888
Combination Product (Y/N)N
Initial Reporter StateAL
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberC-VH-3030
Device Catalogue NumberC-VH-3030
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/23/2023
Supplement Date Received by Manufacturer03/03/2023
Initial Report FDA Received Date02/20/2023
Supplement Report FDA Received Date03/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-