| Model Number |
866470 |
| Medical Device Problem Code |
No Audible Prompt/Feedback (2282)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
|
| |
|
Event or Problem Description
|
|
The customer reported that a message was displayed indicating a speaker operation fault.The device was in use on a patient.There was no report of patient or user harm.A follow-up report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Philips received a complaint on the intellivue patient monitor mx850, indicating displays a speaker malfunction message.No harm to the patient.The field service engineer performed an analysis on the equipment and did not confirm the reported failure.The equipment shows a speaker malfunction message and the sound was coming from the mx800.After restarting the device, it performed as expected and the equipment was released for use.The device was confirmed to be operating per specifications and no failure was identified.The investigation concludes that no further action is required at this time.If additional information is received, the claim file will be reopened.
|
| |
|